PRODUCT

Selectra ProS

Compact fully automated benchtop solution. The Selectra ProS™ benchtop system is the ideal choice for a small or satellite clinical chemistry laboratory seeking a compact fully automated and reliable solution to increase capacity and produce results you can trust.

Together with a broad menu of ready-to-use, liquid-stable reagents, the Selectra ProS provides a fully integrated chemistry solution for smaller laboratories. With random access and a typical throughput of up to 133 tests per hour, the ProS system is the ideal benchtop workhorse for primary, STAT or backup testing needs. Selectra Pro S offers minimal maintenance and effective use of consumables to truly reduce operational costs.

  • Fully-automated clinical chemistry analyzer
  • Typically 133 photometric tests/hour, plus optional dry ISE
  • Best-in-class operator interface
  • Productivity and error reducing capabilities
  • Automated reagent management
  • Annual ProActive Maintenance Kit

Delivering trouble-free results laboratories can trust!


More information available

Selectra ProS International

Image Gallery

Benefits


Processing Center

Single probe streamlines operation

  • Liquid sensor detection
  • Collision protection
  • Precision metering

Optimized test environment ensures accurate results

  • Unique ‘vortex’ mixing
  • Temperature controlled, 37°C

Disposable cuvette rotor with automated onboard wash

  • 10,000 tests / cuvette rotor

Combined reagent and sample rotor with integrated cooling

  • Enables 28 day onboard stability for most reagents
  • 30 reagent positions
  • Ready-to-use, liquid-stable reagents
  • Methodologies referenced to industry standards

Barcoded for easy loading and tracking

  • Bottle barcode loads assay ID, expiration date, lot number, bottle size, test count
  • Test parameters, calibration, and control values also loaded via barcode

Combined reagent and sample rotor with Positive Sample Identification (PSID)

  • Increases productivity and reduces errors
  • Accommodates 25 samples
  • Primary tubes (13 x75 mm) and sample cups

Dry ISE Technology

Patented dry Ion Selective Electrodes (ISE) technology

  • Reduces reagent waste and operating costs
  • Long onboard stability from 3 months – 2 years
  • Longer linear range

Command Center

New best-in-class operator interface with Selectra Touch Pro software

  • Touch screen monitor with next generation software navigation
  • Intuitive color icons guide the operator through daily workflow
  • Rapid access to key operations at a touch

Powerful software capabilities simplify operation

  • Automated sample programming with optional host query LIS interface and PSID
  • Real time tracking to plan and manage on-board inventory
  • 40 configurable result checks enable auto validation
  • Automated programming of assays, calibrators and control values using barcodes
  • Maintenance tasks can be scheduled to run automatically when chemistry analyzer is not in use
  • Powerful reporting and back up functionality

Throughput

  • Up to 133 tests/hou
  • Up to 333 ISE tests/hour

Reagent and Sample System

  • One rotor combining sample and reagent positions
  • Inner rotor ring:
  • 30 refrigerated reagent positions for 10 mL and 25 mL reagent bottles
  • Cooled to 10°C +/- 4°C at normal laboratory conditions
  • All reagent positions can be assigned as R1, R2 and R3.
  • Outer rotor ring:
  • 25 barcode readable positions
  • All positions fit 13×75 mm primary and secondary tubes and pediatric cups
  • All positions can be used for calibrators, controls, normal, pediatric, and STAT samples.

Pipetting System

  • R1 volume 110-399 µl
  • 1000 µl reagent syringe:
  • R2 volume 0-289 µl
  • R3 volume 0-289 µl
  • Programmable in 1 µl steps
  • Sample volume 1-30 µl
  • 100 µl sample syringe:
  • Programmable in 0.1 µl steps
  • Programmable dilution ratios 1:5 up to 1:100 in one-step increments with 3 possible diluents
  • Pre-heated probe with level detection, collision protection, and integrated mixer.

Cuvette Rotor

  • Cost-effective, semi-disposable cuvette rotor with 48 cuvettes
  • 10.000 tests per rotor
  • Typical reagent consumption is 220 µL per test
  • Measurement temperature 37°C, controlled by Peltier elements.

Light Source

  • Quartz-iodine lamp 12V-20W.

Wavelength Range

  • Optical unit with 8-position filter wheel
  • 340, 405, 505, 546, 578, 620, 660, 700 nm standard wavelengths
  • Other wavelengths available on request.

Photometric Range

  • -0.1 to 3.0 Absorbance
  • Resolution 0.001 Abs.

Analytical Modes (Single, Dual and Triple reagent system)

  • Kinetic measurement with linearity check
  • Mono- and bichromatic endpoint measurement with or without bichromatic reagent blank and/or sample blank correction
  • Two-point measurement; with or without slope blank
  • Graphic plot of all measuring points
  • Predilution and automatic reflex dilution as needed
  • Non-linear calibration curves
  • Prozone check for immunology tests
  • Cut-off declaration
  • Calculated tests.

Quality Control

  • Up to 15 controls can be defined, 3 per test
  • Westgard or Basic QC rules
  • Levey-Jennings plots
  • Quality control statistics.

Water Consumption

  • ~950 mL per hour when fully operating.

Standards and Regulations

  • CE -IVDR
  • CB
  • FDA 510(k) cleared
  • UL (selected versions)

Dimensions

  • 83 x 59 x 61 cm (W x H x D)
  • 33 x 23 x 24 in (W x H x D)
  • Weight: 75 kg (165 lbs).

Interface

  • State-of-the-art Host-Query interface available;
  • Host: RS 232 or Ethernet (TCP/IP) through LIS-2A protocol.

Environmental Conditions

  • Temperature: 15 – 32 °C (59 – 90 °F)
  • Humidity: 15 – 85% RH
  • Altitude: up to 2000 m
  • Plumbing: No dedicated water or drain required
  • Electrical:
  • Voltage: 100 – 240 Vac
  • Frequency: 50 – 60 Hz
  • Power (max): 380 VA .

Integrated PC

  • Touch screen navigation
  • Operating system: Windows™ 10 IoT

Options

Barcode Reader

  • Hand-held CCD barcode reader used for reagent identification and automated programming of assays, controls, and calibrators.

Dry ISE Module

  • Patented Solid State Electrode Technology
  • Indirect measurement
  • Dilution 1:14
  • Measures Sodium, Potassium, Chloride and CO2.

Positive Sample Identification

  • Positive Sample Identification (PSID) reads all popular barcodes.

Printer

  • Printer supported by MS Windows™.

ProActive Maintenance Kit

  • Complete parts kit for annual preventive maintenance.

Let us help you

For general inquiries or technical support, please use the link to the right. Click ‘Contact’ to complete a brief online form, and someone will be in touch with you soon.

CONTACT

Navigation

About us

Careers

News

Documentation

Contact

Regions

Clinical Chemistry International

Social Media

Our commitment

Code of Conduct for Suppliers

Terms & Conditions

Privacy Policy